In addition, it is used to detect radiation in the field of wireless LAN, GSM or, for example, microwave radiation. With frequencies up to 3.5 GHz, it can also be used well in the HF range. The three-axis probe saves conversions for the individual axes. Due to its special characteristics, the small, compact radiation measuring device is also suitable for use in industrial environments as well as in the laboratory. Simple, fast and accurate.
| Frequency range | 50 MHz ... 3.5 GHz |
| Sensor type | electric field (E) |
| Measurement | 3-dimensional, isotropic |
| Measuring ranges | 20 mV/m ... 108.0 V/m 53 µA/m ... 286.4 mA/m 1 µw/m² ... 30.93 W/m² 0 µW/cm² ... 3.093 mW/cm² |
| Measuring range selection | automatic |
| Response time | 1 s (until 90% of the final value is reached) |
| Display units | mV/m , V/m , µA/m , mA/m, µW/m² , mW/m² |
| Resolution | 0.1mV/m 0.1 µgA/m 0.01 µgW/m² |
| Measurement error (absolute) (at 1 V/m and 50 MHz) |
±1.0 dB |
| Accuracy (frequency dependent) | ±1.0 dB (50 MHz ... 1.9 GHz) ±2.4 dB (1.9 GHz ... 3.5 GHz) |
| Isotropic deviation | ±1.0 dB (at frequency >50 MHz ) |
| Maximum overrange | 4.2 W/m² (40 V/m) |
| Temperature-related deviation | ±1.5 dB |
| Display | 4-digit LCD |
| Refresh of the display values | every 400 ms |
| Limit | adjustable |
| Alarm | beep when limit is exceeded |
| Calibration factor | adjustable |
| Averaging | adjustable over 4 s ... 15 min |
| Internal measurement memory | 99 records (can be called up again in the display) |
| Measured values that can be displayed | current measured value, maximum value, mean value |
| Supply | 9 V block battery |
| Weight | 350 g / 0.78 lb (including battery) |
N.B. Calibration certificates are issued specifically for the customer and are excluded from the right of return.