Polarimeters with circulation thermostat temperature control P8000-T and P8100-T

Update Terakhir
Negara Asal
Germany
Pembelian minimum
1
Harga
Nego
Kategori
Minta Penawaran
Sharing:

Deskripsi produk

Polarimeters for the laboratory

The polarimeter models P8000-T and P8100-T will allow for a circulation thermostat temperature control in the range between 8 °C and 40 °C when combined with temperature controlled

measurement tubes and a thermostat. That makes them suitable for high-precision measurements in quality control. Especially if a multitude of different temperature controlled measurement

tubes are to be used, the P8000-T and P8100-T are the right choice. We offer a very large selection of this type of tubes.

  • Temperature measurement directly inside the sample
  • Accurate, homogeneous water bath temperature control
  • Large selection of temperature-controlled measurement tubes
  • Quick coupling system for an easy measurement tube change
  • Highest accuracy over the entire measurement range
  • Very short measurement time of approx. 1 s
  • Measurement up to an optical density of 3.0 OD
  • Intuitive operation via touch-screen display
  • Predefined scales (opt. rotation, spec. rotation, int. sugar scale, concentration)
  • Any number of freely definable scales with conversions based on tables or formulas
  • Chemical-resistant materials
  • Compliance with GMP/GLP, 21 CFR Part 11, Pharmacopeia (USP, BP, JP, Ph. Eur.), FDA, ISO, HACCP, OIML, ASTM, ICUMSA, NIST

Model specifications

P8000-T P8100-T
MEASUREMENT DATA
SCALES Optical rotation [°]
Int. sugar scale [°Z]
Concentration [g/100 ml]
Spec. rotation
User-defined
MEASUREMENT RANGE ±90°
±259 °Z
MEASUREMENT ACCURACY ±0.003°
±0.01 °Z
±0.002°
±0.01 °Z
RESOLUTION 0.001°
0.01 °Z
MEASUREMENT PERIOD ±90° Approx. 1 s
LIGHT SOURCE 1 LED with filter
WAVELENGTH 589 nm
TEMPERATURE DATA
TEMPERATURE CONTROL With Circulation Thermostat PT31/PT80 and temperature-controlled measurement tube
TEMPERATURE CONTROL RANGE PT31: 8-35 °C | PT80: 5-80 °C
TEMPERATURE CONTROL ACCURACY ±0.2 °C
TEMPERATURE MEASUREMENT With temperature probe PRT-E or PRT-T
TEMPERATURE MEASUREMENT RANGE 0–99.9 °C
TEMPERATURE MEASUREMENT RESOLUTION 0.1 °C
TEMPERATURE MEASUREMENT ACCURACY ±0.1 °C

 




Hubungi:
PT Bopana Arga Prima Jaya
Jalan Jupiter VI Perum Bintang Metropol
Blok A7 No.16 Bekasi Utara 17122
Jawa Barat, Indonesia
Telp : 021-22163350
Hp : 08119762505 / 081285345374
Email : boney@bapj.co.id / tesa@bapj.co.id

Produk Terkait

                                                                                      PT. Bopana Arga Prima Jaya

Office

Jl. Jupiter VI, Perum Bintang Metropol Blok A7 No.16 Bekasi Utara 17122 , Bekasi
Bekasi , Indonesia

Contact
021 -2216 3350
info@bapj.co.id
Share
apartmentenvelopebookmarkcarttagphone-handsetearthhistorymagnifiercross
× How can I help you?