| SKU | HI83326-01 |
| Name | Pool and Spa Photometer – HI83326 |
| pH Range | Photometer: 6.5 to 8.5 pH pH electrode: -2.00 to 16.00 pH |
| pH Resolution | Photometer: 0.1 pH pH electrode: 0.01 pH |
| pH Accuracy | Photometer: ±0.1 pH pH electrode: ±0.01 pH |
| pH Calibration | Automatic one or two point calibration with one set of standard buffers available (4.01, 6.86, 7.01, 9.18, 10.01) |
| pH Temperature Compensation | Automatic (-5.0 to 100.0 ℃; 23.0 to 212.0 °F); limits reduced based on the pH electrode used |
| pH CAL Check (electrode diagnostics) | clean electrode and check buffer/check probe displayed during calibration |
| pH Method | Photometer: phenol red |
| pH-mV Range | ±1000 mV |
| pH-mV Resolution | 0.1 mV |
| pH-mV Accuracy | ±0.2 mV |
| Absorbance Range | 0.000 to 4.000 Abs |
| Absorbance Resolution | 0.001 Abs |
| Absorbance Accuracy | ±0.003Abs @ 1.000 Abs |
| Alkalinity Range | Freshwater: 0 to 500 mg/L (as CaCO3) |
| Alkalinity Resolution | 1 mg/L |
| Alkalinity Accuracy | ±5 mg/L ±5% of reading |
| Alkalinity Method | Colorimetric method |
| Bromine Range | 0.00 to 8.00 mg/L (as Br2) |
| Bromine Resolution | 0.01 mg/L |
| Bromine Accuracy | ±0.08 mg/L ±3% of reading |
| Bromine Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method. |
| Free Chlorine Range | 0.00 to 5.00 mg/L (as Cl2) |
| Free Chlorine Resolution | 0.01 mg/L |
| Free Chlorine Accuracy | ±0.03 mg/L ±3% of reading |
| Total Chlorine Range | 0.00 to 5.00 mg/L (as Cl2) |
| Total Chlorine Resolution | 0.01 mg/L |
| Total Chlorine Accuracy | ±0.03 mg/L ±3% of reading |
| Chlorine Method | Adaptation of the EPA 330.5 DPD method |
| Copper Range | High range: 0.00 to 5.00 mg/L (as Cu2+) |
| Copper Resolution | 0.01 mg/L |
| Copper Accuracy | High Range ±0.02 mg/L ±4% of reading |
| Copper Method | Adaptation of the EPA bicinchoninate method |
| Cyanuric Acid Range | 0 to 80 mg/L (as CYA) |
| Cyanuric Acid Resolution | 1 mg/L |
| Cyanuric Acid Accuracy | ±1 mg/L ±15% of reading |
| Cyanuric Acid Method | Adaptation of the turbidimetric method |
| Hardness, Calcium Range | 0.00 to 2.70 mg/L (as CaCO3) |
| Hardness, Calcium Resolution | 0.01 mg/L |
| Hardness, Calcium Accuracy | ±0.11 mg/L ±5% of reading |
| Hardness, Calcium Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Calmagite method |
| Iron Range | High Range: 0.00 to 5.00 mg/L (as Fe) |
| Iron Resolution | 0.01 mg/L |
| Iron Accuracy | High Range: ±0.04 mg/L ±2% of reading |
| Iron Method | High Range: Adaptation of Standard Methods for the Examination of Water and Wastewater, 3500-Fe B., Phenanthroline Method |
| Nitrate Range | 0.0 to 30.0 mg/L (as NO3–– N) |
| Nitrate Resolution | 0.1 mg/L |
| Nitrate Accuracy | ±0.5 mg/L ±10% of reading |
| Nitrate Method | Adaptation of the cadmium reduction method |
| Ozone Range | 0.00 to 2.00 mg/L (as O3) |
| Ozone Resolution | 0.01 mg/L |
| Ozone Accuracy | ±0.02 mg/L ±3% of reading |
| Ozone Method | Colorimetric DPD Method |
| Phosphate Range | Freshwater Low Range: 0.00 to 2.50 mg/L (as PO43-) |
| Phosphate Resolution | Freshwater: 0.01 mg/L |
| Phosphate Accuracy | Freshwater Low Range: ±0.04 mg/L ±4% of reading |
| Phosphate Method | Freshwater Low Range: Adaptation of the Ascorbic Acid method |
| Input Channels | 1 pH electrode input and 3 photometer wavelengths |
| pH Electrode | digital pH electrode (not included) |
| Photometer/Colorimeter Light Source | 3 LEDs with 525 nm, 575 nm, and 610 nm narrow band interference filters |
| Photometer/Colorimeter Light Detector | silicon photodetector |
| Bandpass Filter Bandwidth | 8 nm |
| Bandpass Filter Wavelength Accuracy | ±1 nm |
| Cuvette Type | round, 24.6 mm |
| Number of Methods | 128 max. |
| GLP | calibration data for connected pH electrode |
| Display | 128 x 64 pixel LCD with backlight |
| Logging Type | log on demand with user name and sample ID optional input |
| Logging Memory | 1000 readings |
| Connectivity | USB-A host for flash drive; micro-USB-B for power and computer connectivity |
| Power Supply | 5 VDC USB 2.0 power adapter with USB-A to micro-USB-B cable (included) |
| Battery Type/Life | 3.7 VDC Li-polymer rechargeable battery / >500 photometric measurements or 50 hours of continuous pH measurement |
| Environment | 0 to 50.0 ℃ (32 to 122.0 °F); 0 to 95% RH, non-condensing |
| Weight | 1.0 kg (2.2 lbs.) |
| Dimensions | 206 x 177 x 97 mm (8.1 x 7.0 x 3.8 in.) |