The Fluke 59 MAX+ Infrared Thermometer delivers the precision you need to do your job accurately and within your budget. It features a 10:1 distance to spot ratio and precise single laser targeting for accurate and repeatable measurements. Use the Fluke 59 MAX+ to perform non-contact temperature measurements on transformers, motors, pumps, panels, breakers, compressors, duct, and vents. It is extremely easy to use and compact enough to fit in your hand and your tool bag. Designed to withstand a one meter drop, you can count on this lightweight, compact infrared thermometer to work when you need it.
For a 12:1 distance to spot ratio, dual laser targeting check out the Fluke 62MAX+ thermometer.
Other useful features:
Specifications | ||
Temperature Range | -30°C to 500°C (-22°F to 932°F) | |
Accuracy (Calibration geometry with ambient temperature 23°C ±2°C) | ≥ 0°C | ±1.5°C or ±1.5% of reading, whichever is greater |
≥ -10°C to < 0°C: | ±2.0°C | |
< -10°C | ±3.0°C | |
≥ 32°F | ±3.0°F or ±1.5% of reading, whichever is greater | |
≥ 14°F to < 32°F | ±4.0°F | |
< 14°F: | ±6.0°F | |
Response Time (95%) | < 500 ms (95% of reading) | |
Spectral Response | 8 mm to 14 mm | |
Emissivity | 0.10 to 1.00 | |
Distance to Spot Ratio | 10:1 (calculated at 90% energy) | |
Display Resolution | 0.1°C (0.2°F) | |
Repeatability (% of reading) | ±0.8% of reading or ±1.0°C (±2.0°F), whichever is greater |
|
Power | 1 AA IEC LR06 Battery | |
Battery Life | 12 hours with laser and backlight on | |
Weight | 220 g (7.76 oz) | |
Size | 156 x 80 x 50 mm (6.14 x 3.15 x 2 in) | |
Operating Temperature | 0°C to 50°C (32°F to 122°F) | |
Storage Temperature | -20°C to +60°C (-4°F to 140°F), (without battery) | |
Operating Humidity | 10% to 90% RH non-condensing @ 30°C (86°F) | |
Operating Altitude | 2000 meters above mean sea level | |
Storage Altitude | 12,000 meters above mean sea level | |
Drop Test | 1 m | |
Ingress Protection Rating | IP40 per IEC 60529 | |
Vibration and Shock | IEC 68-2-6 2.5 g, 10 to 200 Hz, IEC 68-2-27, 50 g, 11 ms | |
Compliance | EN/IEC 61010-1 | |
Laser Safety | FDA and EN 60825-1 Class II | |
Electromagnetic Compatibility | 61326-1 EN 61326-2 |