• In addition to the facility for precision measurement, test leads for simplified two wire measuring system also supplied as standard accessories.
(unit can be hung from the neck for simplified measurement)
• The latest circuit design permits the instrument to operate with the minimum of influence from earth voltage and earth resistance of auxiliary earth spikes.
• Dust and drip proof. (designed to IEC 60529 IP54)
• Earth resistance value can be read directly from the scale.
• Designed to meet IEC 61010-1 safety standard.
• Capable of measuring earth voltage.
• Small and lightweight. Shock resistant new case material.
• 2mA measuring current permits earth resistance tests without tripping earth leakage current breakers in the circuit under test.
Soft case
*Please check here for the hard case model.
Adapter to enable use of ather test leads. (MODEL 8259)
Earth resistance measurement |
20Ω | 200Ω | 2000Ω | |
---|---|---|---|---|
Measuring range | 0.00 - 1999Ω | |||
Display range | 0.00 - 19.99Ω | 0.0 - 199.9Ω | 0 - 1999Ω | |
Accuracy | ±2%rdg±0.1 Ω | ±2%rdg±3dgt | ||
Earth voltage measurement | ||||
Measuring range | 0 - 200 V AC (50,60Hz) | |||
Display range | 0.0 - 199.9 V | |||
Accuracy | ±1%rdg±4dgt | |||
Overload protection | Earth resistance : 280V AC (10 seconds) Earth voltage : 300V AC (1 minute) |
|||
Applicable Standards | IEC 61010-1 CAT Ⅲ300 V Pollution degree 2 IEC 61010-2-030, IEC 61557-1, -5, IEC 60529 (IP54) |
|||
Power source | R6(AA)(1.5V) × 6 | |||
Dimensions | 105(L) × 158(W) × 70(D) mm (including case lid) | |||
Weight | Approx. 550g (including batteries and case lid) | |||
Accessories | 7095A (Earth resistance test leads [red-20m, yellow-10m, green-5m/1set]) 7127B (Simplified measurement probe), 8032 (Auxiliary earth spikes[2 spikes/1set]), 9121 (shoulder strap), 9084 (Soft case) R6(AA) × 6, Instruction manual |
|||
Optional | 7245A (Precision measurement cord set) 8259 (Adapter for measurement terminal) 9165 (Hard case) |