Fluke 107 measurement test functions
Fluke 107 comes ready to work
The Fluke 107 comes with a TL75 Hard Point™ Test Lead Set. It’s one pair (red, black) of 48 inch (1.22 meter) test leads with right-angle, shrouded banana plugs. The meter also includes a SmartStrap™ intelligent magnetic multi-purpose lanyard that lets you hang the Fluke 107 from any ferrous surface or support the meter at an angle for easy viewing.
Accuracy is available for one year after calibration, operating from 18 °C to 28 °C and relative humidity from 0% to 75%. The format of the accuracy specification is: ± ([% of reading] + [least significant digit])
Precision specifications | ||||
AC voltage (40 Hz to 500 Hz)1 |
Range | Resolution | Accuracy | |
6.000 V 60.00 V 600.0 V |
0.001 V 0.01 V 0.1 V |
1.0 % + 3 | ||
DC voltage | 6.000 V 60.00 V 600.0 V |
0.001 V 0.01 V 0.1 V |
0.5 % + 3 | |
Exchange millivolts | 600.0 mV | 0.1 mV | 3.0 % + 3 | |
Diode test2 | 2.000 V | 0.001 V | 10 % | |
Resistance (ohm) |
400.0 Ω 4.000 kΩ 40.00 kΩ 400.0 kΩ 4.000 MΩ 40.00 MΩ |
0.1 Ω 0.001 kΩ 0.01 kΩ 0.1 kΩ 0.001 MΩ 0.01 MΩ |
0.5 % + 3 0.5 % + 2 0.5 % + 2 0.5 % + 2 0.5 % + 2 1.5% + 3 |
|
Capacitor3 | 50.00 nF 500.0 nF 5.000 μF 50.00 μF 500.0 μF 1000 μF |
0.01 nF 0.1 nF 0.001 μF 0.01 μF 0.1 μF 1 μF |
2 % + 5 2 % + 5 5 % + 5 5 % + 5 5 % + 5 5 % + 5 |
|
Frequency4 Hz (10 Hz – 100 kHz) |
50.00 Hz 500.0 Hz 5.000 kHz 50.00 kHz 100.0 kHz |
0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz 0.1 kHz |
0.1 % + 3 | |
Duty cycle4 | 1 % to 99 % | 0.10 % | 1 % typical case5 | |
AC current (40 Hz - 200 Hz) |
4.000 A 10.00 A |
0.001 A 0.01 A |
1.5 % + 3 | |
DC | 4.000 A 10.00 A |
0.001 A 0.01 A |
1.5 % + 3 | |
|
||||
Features | Overload protection | Input impedance (nominal value) | Common mode rejection ratio | Normal mode rejection ratio |
AC voltage | 600 V 1 | >10 MΩ <100 pF 2 | >60 dB at 50 Hz or 60 Hz at DC current | − |
Exchange millivolts | 600 mV | >1M, <100 pF | >80 dB at 50 Hz or 60 Hz | − |
DC voltage | 600 V 1 | >10 MΩ <100 pF | >100 dB at 50 Hz or 60 Hz at DC current | >60 dB at 50 Hz or 60 Hz |
|
||||
General technical indicators | ||||
The highest voltage applied to any terminal and ground | 600 V | |||
Display (LCD) | 6000 counts, updated three times per second | |||
Type of battery | Two AAA batteries (NEDA 24A, IEC LR03) | |||
Battery life | At least 200 hours | |||
Temperature | ||||
Operating temperature | 0 °C to 40 °C | |||
Storage temperature | -30 °C to 60 °C | |||
Relative humidity | ||||
Working humidity | No condensation below 10 °C ≤ 90 % at 10 °C to 30 °C; ≤ 75 % at 30 °C to 40 °C |
|||
Working humidity, 40 MΩ range | ≤ 80 % at 10 °C to 30 °C; ≤ 70 % at 30 °C to 40 °C | |||
Altitude | ||||
Working elevation | 2000 m | |||
Storage altitude | 12,000 m | |||
Temperature coefficient | 0.1 X (specified accuracy) / °C (<18 °C or >28 °C) | |||
Current input fuse protection | 11A, 1000V fast-melting type, must use Fluke designated parts | |||
Volume (height x width x length) | 142 mm x 69 mm x 28 mm | |||
Weight | 200 g | |||
Protection level | IEC 60529: IP 40 | |||
Safety | IEC 61010-1: 600 V CAT III, pollution degree 2 | |||
Electromagnetic environment | IEC 61326-1: Portable | |||
Electromagnetic compatibility | Applicable only in Korea. Class A equipment (industrial broadcasting and communication equipment)1 | |||
1 This product meets the requirements of industrial (Class A) electromagnetic wave equipment, which should be noted by the seller or user. The device is suitable for use in a work environment, not a home environment. |