Use the programmable pump limit settings to eliminate over-pressurization, and to set “pump to” pressure values. The unique cleanout port design reduces pump failures, and makes the 719 easy to clean without repair. The compact, lightweight form, combined with one-handed calibration, makes this pressure calibrator the ideal tool for busy process professionals.
The Fluke 719 delivers pressure calibration made easy, by a name you know and trust for all your process calibration needs.
Other useful features:
General Specifications | ||
Storage temperature | -30°C to 60°C | |
Operating temperature | -10°C to 55°C | |
Relative humidity | 95% (-10°C to 30°C) | |
75% (30°C to 40°C) | ||
45% (40°C to 50°C) | ||
35% (50°C to 55°C) | ||
Shock | 1 meter drop test | |
Safety | CSA C22.2 No. 1010.1:1992 | |
EMC | EN50082-1:1992 and EN55022:1994 Class B | |
Size (H x W x L) | 60 x 87 x 210 mm | |
(2.19 x 3.41 x 8.28 in) | ||
Weight | 912 g (2 lb) | |
Battery | Two 9 V alkaline batteries | |
Battery life | 12 hours with 12 mA into 500 Ω, 200 pump cycles to 100 psi, 500 pump cycles to 30 psi | |
Warranty | Three-years, one year for pump | |
Specifications | ||
Fluke-719 30G | Range | -12 to 36 psi, -850 mbar to 2.4 bar |
Resolution | 0.001 psi, 0.1 mbar | |
Accuracy | 0.025% range for 6 months 0.035% of range for one (1) year |
|
Fluke-719 100G | Range | -12 to 120 psi, -850 mbar to 8 bar |
Resolution | 0.01 psi, 1 mbar | |
Accuracy | 0.025% range for 6 months 0.035% of range for one (1) year |
|
mA | Range | 0 to 24 mA |
Resolution | 0.001 mA | |
Accuracy | .015% Rdg + 2 counts | |
Max load, 1000 Ω, 750 Ω ohm in HART mode | ||
Loop supply | Range | 24 V DC |
Resolution | N/A | |
Accuracy | ±10% | |
Pump Specifications | ||
Maximum pressure generation | 120 psi (8 BAR) (100 psi model) | |
Pump to | -30 psi in less than 15 seconds1 | |
-100 psi in less than 45 seconds1 | ||
1. With 1 meter of 1/8 inch test hose | ||
Pump cycles between rebuilds | 5000 | |
Total pump life | 20,000 calibration cycle |