The T6-1000 PRO Electrical Tester measures voltage up to 1000 V ac and current up to 200 A ac, all through the open fork and without test lead contact to live voltage. Connect the black lead to ground with the included heavy-duty alligator clip, slide the wire into the open fork and see both voltage and current at the same time. On wires up to AWG 4/0 (120 mm2), carrying as much as 200 A and 1000 V ac. Even with gloves on. It will change how you do your job.
Function | ||
FieldSense ac voltage | Requires test leads | No |
Range | 1000 V | |
Resolution | 1 V | |
Accuracy | ±(3 % + 3 counts) 45 Hz - 66 Hz[1] [2] | |
FieldSense ac current | Requires test leads | No |
Range | 200.0 A | |
Resolution | 0.1 A | |
Accuracy | ±(3 % + 3 counts) 45 Hz - 66 Hz | |
FieldSense frequency (Hz) | Requires test leads | No |
Range | 45 Hz – 66 Hz | |
Resolution | 1 Hz | |
Accuracy | ±(1 % + 2 counts) | |
Volts ac | Requires test leads | Yes |
Range | 1000 V | |
Resolution | 1 V | |
Accuracy | ±(1.5 % + 2 counts) 45 Hz - 66 Hz | |
Volts dc | Requires test leads | Yes |
Range | 1000 V | |
Resolution | 1 V | |
Accuracy | ±(1 % + 2 counts) | |
Resistance | Requires test leads (all ranges) | Yes |
Range | 2000 Ω | |
Resolution | 1 Ω | |
Range | 20.00 kΩ | |
Resolution | 0.01 kΩ | |
Range | 100.0 kΩ | |
Resolution | 0.1 kΩ | |
Accuracy (all ranges) | ±(1 % + 2 counts) |
Accuracy[1]: ± ( [ % of reading] + [number of least significant digits] ). Accuracy is specified for 1 year after calibration, at 18 °C to 28 °C (64 °F to 82 °F) with relative humidity to 90 %. AC conversions are ac-coupled, RMS responding.
Accuracy[2]: FieldSense specified from 16V to 1000V.
General Specifications | |
Calibration | 1-year calibration cycle |
Dimensions | 61 mm x 259 mm x 43 mm |
Weight | 0.35 kg (0.78 lb) |
Fork opening | 17.8 mm (0.7”) |
Temperature Operating: Storage: |
-10 °C to +50 °C (14 °F to 122 °F) -30 °C to +60 °C (-22 °F to +140 °F) |
Altitude Operating: Storage: |
2,000 m (6,562 feet) 10,000 m (32,808 feet) |
Relative Humidity | 0 % to 90 %, 5 °C to 30 °C (41 °F to 86 °F) 0 % to 75 %, 30°C to 40 °C (86 °F to 104 °F) 0 % to 45 %, 40°C to 50 °C (104 °F to 122 °F) |
Battery Type: Life: |
2x AA (IEC LR6) 360 hour continuous, typical is 200 hour using the FieldSense function |
Temperature coefficient | 0.1 x (specified accuracy) / °C for <18°C or >28 °C (< 64.4 °F or > 82.4 °F) |
Safety General: Measurement T6-1000 PRO: |
IEC 61010-1: Pollution Degree 2
IEC 61010-2-032: CAT IV 600 V / CAT III 1000 V |
Electromagnetic Compatibility | IEC/EN 61326-1: Portable |