Fitted with a zirconium dioxide sensor and a nondispersive infrared sensor, the MAT1500 is our number one when it comes to the range of application, performance and precision.
The measurements are completed within only approx. ten seconds. As is the case with the MAT1400, highly precise results can be achieved with single measurements of oxygen
concentrations thanks to the min/max detection even if only a small sample volume is available.
MAT1500 | |
---|---|
MAIN SPECIFICATIONS | |
SCALES | Oxygen Carbon dioxide Nitrogen Any number of user-defined scales |
OXYGEN (O2) | |
MEASUREMENT RANGE | 0–100 vol% |
ACCURACY | ±0.001 vol% (< 1 vol%) ±0.01 vol% (< 6 vol%) ±0.1 vol% (< 35 vol%) ±0.4 vol% (> 35 vol%) |
RESOLUTION | 0.001 vol% |
CARBON DIOXIDE(CO2) | |
MEASUREMENT RANGE | 0–100 vol% |
ACCURACY | ±0.5 vol% |
RESOLUTION | 0.1 Vol.-% |
NITROGEN (N2) | |
ARITHMETIC DETERMINATION | Residual content of the gas mixture |
O2-SENSOR | |
TYPE | Zirconium dioxide sensor ((ZrO2) |
MEASUREMENT PRINCIPLE | Potentiometric measurement |
SERVICE LIFE | Unlimited |
CROSS-SENSITIVITY | Cross-sensitive toward inflammable gases |
DRIFT | Drift-free |
CO2-SENSOR | |
TYPE | Two-channel NDIR sensor |
MEASUREMENT PRINCIPLE | Infrared spectroscopy |
SERVICE LIFE | Unlimited |
CROSS-SENSITIVITY | Not affected by moisture |
DRIFT | Low drift < 1 %/month |
‘
‘
‘
Hubungi:
PT Bopana Arga Prima Jaya
Jalan Jupiter VI Perum Bintang Metropol
Blok A7 No.16 Bekasi Utara 17122
Jawa Barat, Indonesia
Telp : 021-22163350
Hp : 08119762505 / 081285345374
Email : boney@bapj.co.id / tesa@bapj.co.id