HANNA HI83326 Pool and Spa Photometer
HI83326 is a compact, multiparameter photometer for use in pool and spa applications. The meter is one of the most advanced photometers available with an innovative optical design that utilizes a reference detector and focusing lens to eliminate errors from changes in the light source and from imperfections in the glass cuvette. This meter has 12 different programmed methods measuring 11 key water quality parameters and also offers an absorbance measurement mode for performance verification and for users that would like to develop their own concentration versus absorbance curves. Pool and spa parameters include alkalinity, free chlorine, nitrite, and phosphate which are critical to maintaining a healthy system.
To save valuable laboratory benchtop space, the HI83326 doubles as a professional pH meter with its digital pH/temperature electrode input. Now one meter can be used for both photometric and pH measurements
| SKU | HI83326-01 |
| Name | Pool and Spa Photometer – HI83326 |
| pH Range | Photometer: 6.5 to 8.5 pH pH electrode: -2.00 to 16.00 pH |
| pH Resolution | Photometer: 0.1 pH pH electrode: 0.01 pH |
| pH Accuracy | Photometer: ±0.1 pH pH electrode: ±0.01 pH |
| pH Calibration | Automatic one or two point calibration with one set of standard buffers available (4.01, 6.86, 7.01, 9.18, 10.01) |
| pH Temperature Compensation | Automatic (-5.0 to 100.0 ℃; 23.0 to 212.0 °F); limits reduced based on the pH electrode used |
| pH CAL Check (electrode diagnostics) | clean electrode and check buffer/check probe displayed during calibration |
| pH Method | Photometer: phenol red |
| pH-mV Range | ±1000 mV |
| pH-mV Resolution | 0.1 mV |
| pH-mV Accuracy | ±0.2 mV |
| Absorbance Range | 0.000 to 4.000 Abs |
| Absorbance Resolution | 0.001 Abs |
| Absorbance Accuracy | ±0.003Abs @ 1.000 Abs |
| Alkalinity Range | Freshwater: 0 to 500 mg/L (as CaCO3) |
| Alkalinity Resolution | 1 mg/L |
| Alkalinity Accuracy | ±5 mg/L ±5% of reading |
| Alkalinity Method | Colorimetric method |
| Bromine Range | 0.00 to 8.00 mg/L (as Br2) |
| Bromine Resolution | 0.01 mg/L |
| Bromine Accuracy | ±0.08 mg/L ±3% of reading |
| Bromine Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method. |
| Free Chlorine Range | 0.00 to 5.00 mg/L (as Cl2) |
| Free Chlorine Resolution | 0.01 mg/L |
| Free Chlorine Accuracy | ±0.03 mg/L ±3% of reading |
| Total Chlorine Range | 0.00 to 5.00 mg/L (as Cl2) |
| Total Chlorine Resolution | 0.01 mg/L |
| Total Chlorine Accuracy | ±0.03 mg/L ±3% of reading |
| Chlorine Method | Adaptation of the EPA 330.5 DPD method |
| Copper Range | High range: 0.00 to 5.00 mg/L (as Cu2+) |
| Copper Resolution | 0.01 mg/L |
| Copper Accuracy | High Range ±0.02 mg/L ±4% of reading |
| Copper Method | Adaptation of the EPA bicinchoninate method |
| Cyanuric Acid Range | 0 to 80 mg/L (as CYA) |
| Cyanuric Acid Resolution | 1 mg/L |
| Cyanuric Acid Accuracy | ±1 mg/L ±15% of reading |
| Cyanuric Acid Method | Adaptation of the turbidimetric method |
| Hardness, Calcium Range | 0.00 to 2.70 mg/L (as CaCO3) |
| Hardness, Calcium Resolution | 0.01 mg/L |
| Hardness, Calcium Accuracy | ±0.11 mg/L ±5% of reading |
| Hardness, Calcium Method | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Calmagite method |
| Iron Range | High Range: 0.00 to 5.00 mg/L (as Fe) |
| Iron Resolution | 0.01 mg/L |
| Iron Accuracy | High Range: ±0.04 mg/L ±2% of reading |
| Iron Method | High Range: Adaptation of Standard Methods for the Examination of Water and Wastewater, 3500-Fe B., Phenanthroline Method |
| Nitrate Range | 0.0 to 30.0 mg/L (as NO3–– N) |
| Nitrate Resolution | 0.1 mg/L |
| Nitrate Accuracy | ±0.5 mg/L ±10% of reading |
| Nitrate Method | Adaptation of the cadmium reduction method |
| Ozone Range | 0.00 to 2.00 mg/L (as O3) |
| Ozone Resolution | 0.01 mg/L |
| Ozone Accuracy | ±0.02 mg/L ±3% of reading |
| Ozone Method | Colorimetric DPD Method |
| Phosphate Range | Freshwater Low Range: 0.00 to 2.50 mg/L (as PO43-) |
| Phosphate Resolution | Freshwater: 0.01 mg/L |
| Phosphate Accuracy | Freshwater Low Range: ±0.04 mg/L ±4% of reading |
| Phosphate Method | Freshwater Low Range: Adaptation of the Ascorbic Acid method |
| Input Channels | 1 pH electrode input and 3 photometer wavelengths |
| pH Electrode | digital pH electrode (not included) |
| Photometer/Colorimeter Light Source | 3 LEDs with 525 nm, 575 nm, and 610 nm narrow band interference filters |
| Photometer/Colorimeter Light Detector | silicon photodetector |
| Bandpass Filter Bandwidth | 8 nm |
| Bandpass Filter Wavelength Accuracy | ±1 nm |
| Cuvette Type | round, 24.6 mm |
| Number of Methods | 128 max. |
| GLP | calibration data for connected pH electrode |
| Display | 128 x 64 pixel LCD with backlight |
| Logging Type | log on demand with user name and sample ID optional input |
| Logging Memory | 1000 readings |
| Connectivity | USB-A host for flash drive; micro-USB-B for power and computer connectivity |
| Power Supply | 5 VDC USB 2.0 power adapter with USB-A to micro-USB-B cable (included) |
| Battery Type/Life | 3.7 VDC Li-polymer rechargeable battery / >500 photometric measurements or 50 hours of continuous pH measurement |
| Environment | 0 to 50.0 ℃ (32 to 122.0 °F); 0 to 95% RH, non-condensing |
| Weight | 1.0 kg (2.2 lbs.) |
| Dimensions | 206 x 177 x 97 mm (8.1 x 7.0 x 3.8 in.) |
'
'
'