HANNA Inst.
Portable Waterproof pH/EC/TDS Meter (High Range) - HI991301
The HI991301 is designed to measure pH, conductivity (EC), total dissolved solids (TDS), and temperature with a single probe. From setup to calibration all operations are performed with just two buttons. The meter is waterproof and rated for IP 67 conditions.
The HI12883 pre-amplified multiparameter probe features an easy-to-clean sensor and a cloth junction that can be renewed to increase the life of the probe. To ensure against interference from transient electrical noise, a solid-state amplifier is integrated into the HI12883.
User-selectable features include different TDS factors from 0.45 to 1.00 and a range of temperature coefficients (β) from 0.0 to 2.4% for greater consistency and reproducibility. Standardized buffer recognition values are also selectable.
These instruments easily fit in the palm of your hand and the bottom probe connection ensures the electrode cable doesn’t get in your way. The large, multi-level LCD displays the primary reading, temperature, and calibration guides simultaneously.
HI12883 Multiparameter Probe
The HI12883 multiparameter probe includes pH, EC/TDS, and temperature measurement. A solid-state pre-amplifier is integrated into the probe to protect against interference from transient electrical noise.
Amperometric conductivity probe
Built-in temperature sensor
Easy-to-clean
- The sensor is easy to clean and maintain by design
pH sensor with extendable cloth junction
- The HI12883 multiparameter probe features a cloth junction that can be pulled longer, effectively increasing the life of the probe
Features at-a-glance
pH electrode - Preamplified for resistance to electrical noise including that generated by pumps and ballast
EC/TDS sensor - amperometric two pole design
Temperature - NTC sensor for accurate temperature measurement and compensation
Designed to withstand the humidity of a growing environment
pH range is from 0.00 to 14.00 pH
Choice of either a 0.5 or 0.7 conversion factor
0.5 (500 CF) conversion factor is for an EC reading of 10 mS/cm = 5 ppt
0.7 (700 CF) conversion factor is for an EC reading of 10 mS/cm = 7 ppt
- Automatic Temperature Compensation
All readings are compensated for variations in temperature
Temperature displayed in oC or oF along with pH, EC or TDS reading
Meter displays a clock tag that will disappear when the reading has achieved stability
Freezes reading on the display to allow recording of measurement
Meter will automatically shut off if there is not enough power to get an accurate measurement
- Battery % level at startup
When the level is below 5%, the battery symbol on the LCD blinks to indicate a low battery condition.
User selectable automatic shut-off ff to conserve battery life. Options are after 8 minutes, 60 minutes , or disabled.
Specifications :
SKU |
HI991301 |
Product Name |
Portable Waterproof pH/EC/TDS Meter (High Range) - HI991301 |
Quote Required |
Yes |
pH Range |
-2.00 to 16.00 pH; -2.0 to 16.0 pH |
pH Resolution |
0.01 pH; 0.1 pH |
pH Accuracy |
±0.02 pH; ±0.1 pH |
pH Calibration |
automatic, at one or two points with two sets of standard buffers (pH 4.01 / 7.01 / 10.01 or pH 4.01 / 6.86 / 9.18) |
pH Temperature Compensation |
automatic 0 to 50 °C; 32 to 122 °F |
pH CAL Check (electrode diagnostics) |
probe condition indicator |
pH Sensor Check |
yes |
mV Range |
±825 mV (pH‐mV) |
mV Resolution |
1 mV (pH‐mV) |
mV Accuracy |
±1 mV (pH‐mV) |
EC Range |
0.00 to 20.00 mS/cm |
EC Resolution |
0.01 mS/cm |
EC Accuracy |
±2% F.S. |
EC Calibration |
automatic, one-point at 12.88 mS/cm |
TDS Range |
0.00 to 10.00 ppt (g/L) |
TDS Resolution |
0.01 ppt (g/L) |
TDS Accuracy |
±2% F.S. |
TDS Calibration |
automatic |
EC/TDS Temperature Compensation |
automatic |
Temperature Correction Coefficient |
β selectable from 0.0 to 2.4 %/ °C with 0.1 increments |
EC to TDS Conversion Factor |
selectable from 0.45 to 1.00 with 0.01 increments |
Temperature Range |
‐5.0 to 105.0 °C; 23.0 to 221.0 °F |
Temperature Resolution |
0.1°C; 0.1°F |
Temperature Accuracy |
±0.5°C; ±1°F |
pH/EC Probe |
HI12883 PVC body, pre-amplified multiparameter probe with internal temperature sensor, quick connect DIN connector and 1m cable |
Automatic Shut-Off |
user selectable |
Battery Type/Life |
1.5V (3) AAA / approximately 600 hours of continuous use |
IP Protection |
IP67 |
Environment |
0 to 50°C (32 to 122°F); RH max 100% |
Dimensions |
154 x 63 x 30 mm (6.1 x 2.5 x 1.2”) |
Weight |
196 g (6.91 oz.) |
Ordering Information |
HI991301 is supplied complete with HI12883 pH/EC/temperature probe, pH 4.01 calibration buffer sachet (1), pH 7.01 calibration buffer solution sachet (1), HI70030 12.88 mS/cm calibration solution sachet (1), HI70038 6.44 ppt calibration solution sachet (1), HI700601 electrode cleaning solution sachet (2), 100 mL beaker (1), batteries, calibration certificates for both meter and probe, instruction manual, and rugged thermoformed carrying case. |