Regardless of a simple or demanding viscosity measurement - IKA ROTAVISC lo-vi HELI Complete determines the viscosity of low-viscosity liquids in all areas of application ranging from the laboratory to quality control.
The scope of delivery includes a standard spindle set (SP 1-SP 4), a protective bracket, temperature sensor and HELISTAND stand incl. T-spindle set.
Easy to operate
The generous 4.3″ TFT display allows for an intuitive menu guidance. A digital spirit level also supports the correct set-up of the start-up condition. The simple ramp function simplifies repetitive tasks.
Highest measurement accuracy
The level of measurement accuracy of the ROTAVISC rotational viscometer for both Newtonian and non-Newtonian fluids is +/- 1 % of the measuring range. The reproducibility is +/- 0.2 %.
Stepless speed
ROTAVISC is a viscometer, which offers stepless speed adjustment.
Viscosity measurement in motion
The motorized IKA HELISTAND moves the mounted viscometer and its rotating body slowly and uniformly up and down in the sample
IKA ROTAVISC lo-vi HELI Complete
Viscosity measuring range: 15 - 6.000.000 mPas
Scope of delivery
Viscosity Measuring Range | 6000000 mPas |
Viscosity Accuracy (FSR) | 1 % |
Viscosity Repeatibility (FSR) | 0.2 % |
Spring torque | 0.0673 mNm |
Guard rail | lo-vi |
Measuring spindle series | SP set-1 |
Motor rating output | 4.8 W |
Overload protection | yes |
Direction of rotation | right |
Display | TFT |
Speed display | TFT |
Speed range | 0.01 - 200 rpm |
Setting accuracy speed | ±0.01 rpm |
Speed adjustment | TFT |
Torque display | yes |
Torque measurement | yes |
Timer | yes |
Timer display | TFT |
Time setting range | 0.017 - 6000 min |
Temperature measurement resolution | 0.1 K |
Working temperature display | TFT |
Connection for ext. temperature sensor | PT 100 |
Graph function | yes |
Operating mode | timer and continuous operation |
Calibration option | yes |
Touch function | yes |
Pemitted density | 9999 kg/dm3 |
Working temperature | -100 - 300 °C |
Fastening on stand | extension arm |
Support rod diameter (with integrated fastening on stand) | 16 mm |
Telescope stand stroke | 142 mm |
Plug-in coupling (Ø) | 12 mm |
Basic container volume | 600 ml |
Stand | HELISTAND |
Stroke max. | 61 mm |
Diameter | 16 mm |
Dynamic load | 3.2 kg |
Dimensions (W x H x D) | 351 x 629 x 372 mm |
Weight | 7.1 kg |
Permissible ambient temperature | 5 - 40 °C |
Permissible relative humidity | 80 % |
Protection class according to DIN EN 60529 | IP 40 |
RS 232 interface | yes |
USB interface | yes |
Voltage | 100 - 240 V |
Frequency | 50/60 Hz |
Power input | 24 W |
Power input standby | 0.06 W |
DC Voltage | 24 V= |
Current consumption | 1000 mA |