Safety at all times: High-end functions for incubation of cell cultures, bacteria cultures or tissue cultures.

Temperature

Working-temperature range : 5 °C above ambient temperature up to +50 °C
Standard sterilisation programme: 60 minutes at 180°C (without removing the sensors)
Setting temperature range : +18 to +50 °C
Setting accuracy temperature : 0.1 °C
Temperature sensor : 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Temperature variation in chamber : at + 37 °C +/- 0.3 K
Temperature fluctuations with time : at 37 °C +/- 0.1 K

Humidity

Humidity control (standard) : Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93 % rh +/- 2.5 %
Setting accuracy humidity : 0.5 % rh
Setting range active humidity control (with option K7) : 40 to 97 % rh and rh-Off

Control of standard components

CO2 control : Digital electronic CO2 control with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation

Adjustment range

 CO2 :

 

0 to 20 % CO2

Setting accuracy

CO2 :

 

0,1%

Variation in time

CO2  :

 

+/- 0.2 % CO2

Adjustment range

O2 :

 

1 to 20 % O2

Setting accuracy

O2 :

 

0.1 % O2

Control technology

ControlCOCKPIT : TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Function SetpointWAIT : the process time does not start until the set temperature is reached
Adjustable parameters : temperature (Celsius or Fahrenheit), CO2, programme time, time zones, summertime/wintertime

Communication

Interface: Ethernet LAN, USB
Documentation: programme stored in case of power failure
Programming: AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port

Safety

AutoSAFETY : additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating function is switched off in case of overtemperature, cooling function in case of undertemperature
Autodiagnostic system : integral fault diagnostics for temperature and CO2
Alarm : visual and acoustic

Heating concept

6 sides : large-area multi-function heating system on four sides with additional door and back heating to avoid condensation

Standard equipment

Works calibration certificate: Works calibration certificate at +37 °C, 5 % CO2 and 90 % rh (requires option K7); standard equipment for units with active humidity control
Works calibration certificate: Works calibration certificate at +37 °C, 5 % CO2, 90 % rh and 10 % O2 (requires option K7 and option T6); standard equipment for units with O2 control
Works calibration certificate: incl. works calibration certificate (measuring point chamber centre) at +37°C, 5 % CO2 for standard units
Door : fully insulated stainless steel door with 2-point locking (compression door lock)
Door : inner glass door with opening (8 mm Ø) to take gas sample
Standard equipment : CO2 connection set: hose with coupling and clamp
Standard accessories : Membrane filter (in order to remove impurities and pollutants, all incoming gases pass through a membrane filter before they reach the chamber)
Internals : 1 perforated stainless steel shelf/shelves
Internals : 1 stainless steel water dish

Stainless steel interior

Dimensions : w(A) x h(B) x d(C): 400 x 425 x 330 mm (d less 35 mm for fan)
Interior : material 1.4301 (ASTM 304), corrosion resistant
Volume : 56 l
Max. number of internals: 5
Max. loading of chamber : 75 kg
Max. loading per internal : 15 kg

Textured stainless steel casing

Dimensions : w(D) x h(E) x d(F): 559 x 795 x 521 mm (d +56mm door handle)
Housing : rear zinc-plated steel

Electrical data

Voltage Electrical load : 230 V, 50/60 Hz
approx. 1100 W
Voltage Electrical load : 115 V, 50/60 Hz
approx. 1100 W

Ambient conditions

Set Up : The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm.
Ambient temperature : 10 °C to 35 °C
Humidity rh : max. 70 %, non-condensing
Altitude of installation : max. 2,000 m above sea level
Overvoltage category : II
Pollution degree : 2

Packing/shipping data

Transport information : The appliances must be transported upright
Customs tariff number : 8419 8998
Country of origin : Federal Republic of Germany
WEEE-Reg.-No. : DE 66812464
Dimensions approx incl. carton : w x h x d: 730 x 950 x 640 mm
Net weight  : approx. 55 kg
Gross weight carton : approx. 74 kg

Standard units are safety-approved and bear the test marks

CE LGA
VDE
UL
CATEGORY PRODUK
Description
 ICO50medICO50med                                                       Details
ICO105medICO105med                                                     Details
ICO150medICO150med                                                       Details
ICO240medICO240med                                                       Details



Hubungi:
PT Bopana Arga Prima Jaya
Jalan Jupiter VI Perum Bintang Metropol
Blok A7 No.16 Bekasi Utara 17122
Jawa Barat, Indonesia
Telp : 021-22163350
Hp : 08119762505 / 081285345374
Email : boney@bapj.co.id / tesa@bapj.co.id

Produk Terkait

                                                                                      PT. Bopana Arga Prima Jaya

Office

Jl. Jupiter VI, Perum Bintang Metropol Blok A7 No.16 Bekasi Utara 17122 , Bekasi
Bekasi , Indonesia

Contact
021 -2216 3350
info@bapj.co.id
Share
apartmentenvelopebookmarkcarttagphone-handsetearthhistorymagnifiercross
× How can I help you?