PRODUCT DESCRIPTION

The durable and affordable CB-S series incubators are suitable for routine cell cultivation applications in the field of cell production, particularly in academic settings. They are contamination-proof due to 180°C hot air sterilization and reliably pH-stable thanks to the drift-free infrared CO₂ measurements conducted in real time.

IMPORTANT CHARACTERISTICS

  • Temperature range:

    +6°C above ambient temperature to +50°C

  • Humidity range:

    90 to 95% RH

  • CO₂ control range:

    0 to 20 vol.% CO₂

  • Tried-and-tested anti-contamination concept with 180°C hot air sterilization
  • Humidification system with condensation protection
  • CO2 gas mixing jet with Venturi effect
  • CO2 sensor with infrared technology
  • LCD to display temperature and CO2 along with additional information and alarms
  • Internal data logger, measured values can be read out in open format via USB
  • Tightly fitted inner door made of safety glass
  • Seamless stainless steel inner chamber
  • Perforated shelves, stainless steel
  • Troubleshooting system with visual and acoustic alarms
  • Interfaces:

    Ethernet, USB, zero-voltage alarm contact

  • Units can be stacked with stacking adapters; the CB-S is suitable for stacking with CB and CBF models of the same size

Technical specifications

Designation CBS260-230V CBS260UL-120V
Article Number 9640-0003 9640-0004
Option model Standard Standard

Designation CBS260-230V CBS260UL-120V
Temperature range +6 °C above ambient temperature to 50 °C +6 °C above ambient temperature to 50 °C
Temperature uniformity at 37°C 0.4 ± K 0.4 ± K
Temperature fluctuation at 37°C 0.1 ± K 0.1 ± K
Recovery time after door was opened for 30 s at 37°C 8 min 8 min

Designation CBS260-230V CBS260UL-120V
Humidity range 90 ...95 % RH 90 ...95 % RH

Designation CBS260-230V CBS260UL-120V
CO₂ range 0 ...20 Vol.-% CO2 0 ...20 Vol.-% CO2
CO₂ measuring technology IR IR
CO₂ recovery time after door was opened for 30 s at 5 vol. % CO₂ 5 min 5 min

Designation CBS260-230V CBS260UL-120V
Rated Voltage 200...230 V 100...120 V
Power frequency 50/60 Hz 50/60 Hz
Nominal power 1,5 kW 1,5 kW
Unit fuse 10,0 A 16,0 A
Phase (Nominal voltage) 1~ 1~

Designation CBS260-230V CBS260UL-120V
Interior volume 267 L 267 L
Net weight of the unit (empty) 116 kg 116 kg
Load per rack 10 kg 10 kg
Permitted load 40 kg 40 kg
Wall clearance back 100 mm 100 mm
Wall clearance sidewise 50 mm 50 mm

Designation CBS260-230V CBS260UL-120V
Width net 740 mm 740 mm
Height net 1,020 mm 1,020 mm
Depth net 785 mm  785 mm

Designation CBS260-230V CBS260UL-120V
Interior width 620 mm 620 mm
Interior height 620 mm 620 mm
Interior depth 575 mm 575 mm
Inner doors 1 1
Unit doors 1 1

Designation CBS260-230V CBS260UL-120V
Sound-pressure level 41 dB(A) 41 dB(A)
Energy consumption at 37°C 55 Wh/h 55 Wh/h

Designation CBS260-230V CBS260UL-120V
Number of shelves (std./max.) 2/8 2/8