A BINDER material test chamber with mechanical convection of the FP series provides reliably short drying times and particularly fast heating – even for chambers under full loads.
+10°C above ambient temperature up to +300°C
Ethernet
Designation | FP115-230V | FP115UL-120V |
Article Number | 9010-0364 | 9010-0368 |
Option model | Standard | Standard |
Designation | FP115-230V | FP115UL-120V |
Temperature range | +10 °C above ambient temperature to 300 °C | +10 °C above ambient temperature to 300 °C |
Temperature uniformity at 150°C | 1.4 ± K | 1.4 ± K |
Temperature fluctuation at 150°C | 0.3 ± K | 0.3 ± K |
Heating up time to 150°C | 19 min | 19 min |
Recovery time after door was opened for 30 s at 150°C | 5 min | 5 min |
Designation | FP115-230V | FP115UL-120V |
Air change rate at 100°C | 32 x/h | 32 x/h |
Designation | FP115-230V | FP115UL-120V |
Rated Voltage | 230 V | 120 V |
Power frequency | 50/60 Hz | 60 Hz |
Nominal power | 1,4 kW | 1,4 kW |
Unit fuse | 10,0 A | 16,0 A |
Phase (Nominal voltage) | 1~ | 1~ |
Designation | FP115-230V | FP115UL-120V |
Interior volume | 116 L | 116 L |
Net weight of the unit (empty) | 55 kg | 55 kg |
Load per rack | 30 kg | 30 kg |
Permitted load | 150 kg | 150 kg |
Wall clearance back | 160 mm | 160 mm |
Wall clearance sidewise | 100 mm | 100 mm |
Designation | FP115-230V | FP115UL-120V |
Width net | 732 mm | 732 mm |
Height net | 732 mm | 732 mm |
Depth net | 636 mm | 636 mm |
Designation | FP115-230V | FP115UL-120V |
Interior width | 550 mm | 550 mm |
Interior height | 550 mm | 550 mm |
Interior depth | 383 mm | 383 mm |
Unit doors | 1 | 1 |
Designation | FP115-230V | FP115UL-120V |
Sound-pressure level | 43 dB(A) | 43 dB(A) |
Energy consumption at 100°C | 230 Wh/h | 230 Wh/h |
Energy consumption at 150°C | 370 Wh/h | 370 Wh/h |
Designation | FP115-230V | FP115UL-120V |
Number of shelves (std./max.) | 2/5 | 2/5 |
Drying and heating chambers ED | ED-S | FD | FD-S | FED | FP | M series |
PDF, 2.21 MB |
Drying and heating chambers (FP series) - Drying chambers, constant climate chambers, dynamic climate chambers and cooling incubators test the service life of MinebeaMitsumi spindle motors MinebeaMitsumi Technology Center Europe GmbH |
PDF, 479.77 KB |
FP series |
PDF, 7.75 MB |
Data Sheet Model FP 115 |
PDF, 794.58 KB |