JAN code : 4974425800346
Calibration : Available
JCSS Accredited Calibration : Not available
SK-8300 is an infrared thermometer with a wide measuring range from –30°C to 1550°C. Suitable for extremely high temperature applications.
Distance-to-spot ratio is 50:1 thus it enables measurement of a narrower area. Emissivity is adjustable between 0.10 and 1.00 in steps of 0.01.
■ One of the best advantages of a contactless infrared thermometer is its quick response. With SK-8300 you can easily and instantly measure surface temperature in one hand.
■ Applicable for high temperature, and a wide range of temperature from −30 to 1550°C. Using a single-point laser marker helps you identify the target at a glance.
■ D:S = 50:1 (D is distance to the object, S is a diameter of a spot circle)
Emissivity is adjustable in steps of 0.01.
■ Hi/Lo alarm notifies you the set temperature by sound.
■ Memorizes up to 10 measured data.
■ Displays max./min. values, average, and temperature difference during measurement (ΔT)
■ Automatic measurement enables continuous measuring without pressing the trigger.
■ LCD backlight improves readability in a dark place.
■ Turns off the power automatically if the device is not in use, to prevent from running out the battery even when you forget to turn it off.
■ Fixable to a tripod.
■ Certified for PSC, a standard based on Japan's Consumer Product Safety Act, as a class II portable laser device.
■ Caution! This device is for industrial use, not for clinical use.
■ What is the distance-to-spot ratio?
A distance-to-spot ratio, which is the ratio of the Distance between the object and the thermometer to the size of the Spot, is preset at D:S = 50:1 for SK-8300.
For example, if the distance is 500 mm, a spot circle of diameter 10 mm can be measured. Make sure that the object is larger than the circle.
The spot circle is defined as an area to receive 90% of the emitted energy from the object.
■ Warning: Do not look into the laser marker source or point it on eyes as it may damage eyes.
SK-8300 conforms to the Class II in the Standard for Safe Use of Laser Products (JIS C6802) which is defined as follows: A visible light (wavelength between
400 and 700 nm) with an output level less than 1 mW that a human's protective response can resist.
PSC mark is labeled on the device to prove its quality based on Japan's Consumer Product Safety Act.
Cat. No. | 8268-00 |
Product name | Infrared Thermometer |
Model name | SK-8300 |
Measuring range | −30 to 1550°C |
Accuracy | ±3°C (−30.0 to −20.1°C) ±2°C (−20.0 to 100.0°C) ±2%rdg (other) Under blackbody calibration with ambient temperature 23±5°C, emissivity 0.95 |
Resolution | 0.1°C |
Repeatability | ±0.1°C |
Response time | 90% in 0.5 seconds |
Emissivity | Adjustable between 0.10 and 1.00 in steps of 0.01 |
Distance coefficient | D:S = 50:1 (where D is distance, S is diameter of spot circle) |
Detector | Thermopile |
Measured wavelength | 8 to 14 μm |
Laser marker | Red single-point laser Source wavelength: 650 nm Output: Less than 1 mW Certified for PSC, a standard based on Japan's Consumer Product Safety Act, as a class II portable laser device. |
Operation ambient | 0 to 50°C, less than 85%rh without condensing |
Power | 9-volt battery (006P) x 1 |
Battery life | Approx. 8 hours if both laser and backlight are on |
Material | ABS resin (main body), Si (lens) |
Dimensions | 51 (W) x 200 (H) x 166 (D) mm |
Weight | 280 g including a battery |
Standard accessories | Hand strap x1, hard case x1, 9-volt alkaline battery x1, instruction manual |
‘
‘
‘
Hubungi:
PT Bopana Arga Prima Jaya
Jalan Jupiter VI Perum Bintang Metropol
Blok A7 No.16 Bekasi Utara 17122
Jawa Barat, Indonesia
Telp : 021-22163350
Hp : 08119762505 / 081285345374
Email : boney@bapj.co.id / tesa@gmail.com