JAN code : 4974425801404
Calibration : May be available depends on probes
JCSS Accredited Calibration : May be available depends on probes
SK-L751 is a 2-channel temperature logger capable of recording 16000 data. All procedures of configuration, monitoring, downloading, and analysis are finished via PC. °C/°F switchable.
A sensor probe must be purchased in addition. Photo describes an example with a plug-in probe on ch. 1 and a cord probe on ch. 2.
■ Two sensor probes connectable
Choose the best probe for your purpose, as we offer a variety of interchangeable probes. Note that one probe must be purchased separately at least.
■ IP65 dustproof/waterproof
SK-L751 conforms IP65 if the following conditions are met;
- Either one or two of cord probes SK-L751-2/21/22/23/24/25/31/32 are connected
- AC adapter is not used
■ Memorizes up to 16000 data
Available to record data for 4 hours with a logging interval of 1 second and for as long as 2 years and 9 months with an interval of 90 minutes.
* Maximum number of data x logging interval = maximum logging time. E.g., 16000 x 5 sec. = 22.2 hours
■ No data loss even if battery runs out
Data is securely stored in the memory to avoid losing even if there is a power failure.
■ AC/DC input compatible
This product functions well with batteries, however, it is recommended to use AC inlets for measurement in a long duration.
When the AC adapter is connected while the batteries are installed, power is supplied from AC inlet preferentially, and it switches to using the batteries in case of a problem such as blackouts.
■ Data can be copied to PC and micro SD
Logged data can not only be downloaded to a PC but also be copied to a micro SD card (sold separately). Copying can be done even when logging.
■ Dedicated data analysis software
Logged data can be fully utilized such as by graph plotting and real-time display with the software downloadable from our website. It can be also exported in CSV format to open it in a spreadsheet system.
■ Four logging mode options selectable
- Once (default)
Keeps logging until the memory becomes full. Stops logging when 16000 data are recorded.
- Repeatedly (overwriting)
Overwrites the existing data from the beginning once the 16000 data are recorded. Ideal for the case you only need to store data of the latest period.
- Ends at (limited amount)
Stops logging when the number of data reaches the limit you set. Ideal for limiting the amount of data to be collected.
- By page (partitioning)
Partition the memory into four blocks, each having the capacity to store 4000 data. Once you stop logging and restart it the data is stored in the next block, thus it is convenient to sort and manage the data of different time or locations.
■ Fourteen different logging intervals
Available logging intervals: 1, 2, 5, 10, 15, 30 seconds and 1, 2, 5, 10, 15, 30, 60, 90 minutes.
■ Scheduled logging
Available to preset the time to start logging automatically.
■ Key lock function
Prevents misoperation while in logging or in standby for preset start.
■ Warning display for upper and lower thresholds
An alarm light is lit and Hi/Lo indicator blinks if the temperature exceeds or falls below the preset thresholds. The light and the indicator turn off when the measured value falls back within the range of the set values.
■ ID number for each device
A 6-digit ID number can be set for easy identification when multiple loggers are used.
■ °C/°F switchable
■ Features of analysis software
Main window
Configures the settings for the main unit, downloads measured data, and analyzes it.
- Both the logged data in the main unit and the data copied to a micro SD (sold separately) can be downloaded.
- Up to ten units can be controlled using a USB hub.
- Imported data can be zoomed in and out, plotted, exported, printed, and analyzed in various ways such as partial analysis and integral analysis.
- Exporting the data in CSV format enables other spreadsheet software to edit it.
Real-time monitoring
Displays measured data in real-time chart when the main unit and the PC are connected with a communication cable.
- Updates current value, maximum, minimum, average, WBGT value, and alarm information in real time.
- Up to ten units can be monitored simultaneously using a USB hub if all units are the same model. These units must be all SK-L751 (temperature type) or all SK-L754 (temperature/humidity type).
- Graph plotting and exporting are available during monitoring.
- Alarm for Hi/Lo thresholds, WBGT, and battery shortage are available. It has also an option to send email to a preset address when alarm is activated.
* Windows PC only
■ Requirements for analysis software
"Datalogger for Windows"
OS | Windows 8.1, 10, 11 (32/64 bit) with latest service pack installed |
Memory | 4 GB or more (32 bit), 8 GB or more (64 bit) |
HDD | 4 GB or more free space |
Other | USB port, internet connection, and email address are required. |
The latest version of the software can be downloaded from our website.
Cat. No. | 8800-00 |
Product name | Datalogger (Temperature Measurement) |
Model name | SK-L751 |
Applicable probes | Plug-in Type Probe SK-L751-1, With-sensor-cord Type Probe SK-L751-2, etc. |
Measuring range | −40 to 105°C / −40 to 221°F at maximum but depends on probes |
Measuring accuracy | Refer to specifications of probes |
Display accuracy | ±0.2°C / 0.4°F (0 to 40°C / 32 to 104°F) ±0.3°C / 0.6°F (other) |
Resolution | 0.1°C / 0.1°F |
Operation ambient | −10 to 60°C, less than 85%rh without condensing |
Refresh time | 1 second |
Indicated articles | Alarm, four types of logging options, logging/standby/real-time communication mode indicator, memory usage bar, micro SD in/out, low battery mark, key lock indicator, error messages, and other configured values |
Memory capacity | 16000 data per channel |
Logging interval | 14 patterns: 1, 2, 5, 10, 15, 30 sec., 1, 2, 5, 10, 15, 30, 60, or 90 min. |
External memory | Micro SD/SDHC up to 32 GB formatted to FAT32 * NTFS format is not available. |
Dimensions | 114 (W) x 80 (H) x 35 (D) mm except protrusion |
Weight | 232 g including batteries |
Power | AA battery x2 or AC adapter sold separately |
Battery life | Approx. 500 hours on measurement mode, or until 16000 data are recorded on logging mode if display is turned off |
Waterproofness | IP65 dustproof/waterproof when connected with probes of SK-L751-2, -21, -22, -31, or -32 |
Standard accessories | 1.5-m communication cable x1, AA alkaline battery x2, instruction manual |
‘
‘
‘
Hubungi:
PT Bopana Arga Prima Jaya
Jalan Jupiter VI Perum Bintang Metropol
Blok A7 No.16 Bekasi Utara 17122
Jawa Barat, Indonesia
Telp : 021-22163350
Hp : 08119762505 / 081285345374
Email : boney@bapj.co.id / tesa@bapj.co.id