Generation unit can be used for various simulation tests instead of sensor output
U8794 | For the MR8740-50 |
---|
Output terminal | Number of channels: 8 (each channel is isolated), Connector: 25-pin D-sub Max. rated voltage to earth: 25 V |
|||||
---|---|---|---|---|---|---|
Output items | DC voltage, DC current, resistance (simulated output) | |||||
Output range | DC voltage: -0.100 0 V to +5.300 0 V (setting resolution: 0.1 mV) DC current: 5 mA range: -5.000 0 mA to +5.000 0 mA, Setting resolution: 0.1 μA 1 mA range: -1.000 00 mA to +1.000 00 mA, Setting resolution: 0.01 μA 250 μA range: -250. 00 μA to +250.00 μA, Setting resolution: 0.01 μA 50 μA range: -50. 000 μA to +50. 000 μA, Setting resolution: 0.001 μA Resistance: 10 Ω to 1 MΩ, Setting resolution: 6 digits |
|||||
Output accuracy | DC voltage: 5 V range, ±0.035% of setting ± 800 μV DC current: 5 mA range: ±0.050% of setting ± 4.0 μA 1 mA range: ±0.050% of setting ± 800 nA 250 μA range: ±0.050% of setting ± 200 nA 50 μA range: ±0.050% of setting ± 40 nA |
|||||
Other | Self diagnostic, switch output terminals, estimate target connection, cancel offset | |||||
Dimensions and mass | 106 mm (4.17 in) W × 19.8 mm (0.78 in) H × 196.5 mm (7.74 in) D, 280 g (9.9 oz) | |||||
Included accessories | None |